CVE-2014-5433

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.
References
Link Resource
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01 Mitigation US Government Resource Third Party Advisory
Configurations

Configuration 1 (hide)

AND
cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:6.05:*:*:*:*:*:*:*
OR cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*
cpe:2.3:h:baxter:wireless_battery_module:16:*:*:*:*:*:*:*

History

No history.

Information

Published : 2019-03-26 16:29

Updated : 2023-12-10 12:59


NVD link : CVE-2014-5433

Mitre link : CVE-2014-5433

CVE.ORG link : CVE-2014-5433


JSON object : View

Products Affected

baxter

  • sigma_spectrum_infusion_system_firmware
  • wireless_battery_module
  • sigma_spectrum_infusion_system
CWE
CWE-255

Credentials Management Errors

CWE-312

Cleartext Storage of Sensitive Information