CVE-2020-16224

In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.
References
Configurations

Configuration 1 (hide)

OR cpe:2.3:a:philips:patient_information_center_ix:c.02:*:*:*:*:*:*:*
cpe:2.3:a:philips:patient_information_center_ix:c.03:*:*:*:*:*:*:*

History

12 Dec 2023, 21:15

Type Values Removed Values Added
References
  • () https://www.philips.com/productsecurity -
Summary (en) Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart. (en) In Patient Information Center iX (PICiX) Versions C.02, C.03, the software parses a formatted message or structure but does not handle or incorrectly handles a length field that is inconsistent with the actual length of the associated data, causing the application on the surveillance station to restart.

Information

Published : 2020-09-11 14:15

Updated : 2023-12-12 21:15


NVD link : CVE-2020-16224

Mitre link : CVE-2020-16224

CVE.ORG link : CVE-2020-16224


JSON object : View

Products Affected

philips

  • patient_information_center_ix
CWE
CWE-130

Improper Handling of Length Parameter Inconsistency