Vulnerabilities (CVE)

Filtered by vendor Philips Subscribe
Total 101 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2018-5466 1 Philips 1 Intellispace Portal 2023-12-10 5.0 MEDIUM 7.5 HIGH
Philips IntelliSpace Portal all versions of 8.0.x, and 7.0.x have a self-signed SSL certificate vulnerability this could allow an attacker to gain unauthorized access to resources and information.
CVE-2017-9654 1 Philips 1 Dosewise 2023-12-10 4.0 MEDIUM 8.8 HIGH
The Philips DoseWise Portal web-based application versions 1.1.7.333 and 2.1.1.3069 stores login credentials in clear text within backend system files. CVSS v3 base score: 6.5, CVSS vector string: AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:N/A:N.
CVE-2018-5468 1 Philips 1 Intellispace Portal 2023-12-10 7.5 HIGH 9.8 CRITICAL
Philips Intellispace Portal all versions 7.0.x and 8.0.x have a remote desktop access vulnerability that could allow an attacker to gain unauthorized access and in some cases escalate their level of privilege or execute arbitrary code
CVE-2018-8857 1 Philips 8 Brilliance Ct Big Bore, Brilliance Ct Big Bore Firmware, Brilliance 64 and 5 more 2023-12-10 7.2 HIGH 7.8 HIGH
Philips Brilliance CT software (Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior) contains fixed credentials, such as a password or cryptographic key, which it uses for its own inbound authentication, outbound communication to external components, or encryption of internal data. An attacker could compromise these credentials and gain access to the system.
CVE-2018-5458 1 Philips 1 Intellispace Portal 2023-12-10 5.0 MEDIUM 7.5 HIGH
Philips IntelliSpace Portal all versions of 8.0.x, and 7.0.x have a vulnerability using SSL legacy encryption that could allow an attacker to gain unauthorized access to resources and information.
CVE-2018-8853 1 Philips 8 Brilliance Ct Big Bore, Brilliance Ct Big Bore Firmware, Brilliance 64 and 5 more 2023-12-10 7.2 HIGH 8.8 HIGH
Philips Brilliance CT devices operate user functions from within a contained kiosk in a Microsoft Windows operating system. Windows boots by default with elevated Windows privileges, enabling a kiosk application, user, or an attacker to potentially attain unauthorized elevated privileges in Brilliance 64 version 2.6.2 and prior, Brilliance iCT versions 4.1.6 and prior, Brillance iCT SP versions 3.2.4 and prior, and Brilliance CT Big Bore 2.3.5 and prior. Also, attackers may gain access to unauthorized resources from the underlying Windows operating system.
CVE-2017-9657 1 Philips 2 Intellivue Mx40, Intellivue Mx40 Firmware 2023-12-10 3.3 LOW 6.5 MEDIUM
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
CVE-2018-5454 1 Philips 1 Intellispace Portal 2023-12-10 6.8 MEDIUM 8.1 HIGH
Philips IntelliSpace Portal all versions of 8.0.x, and 7.0.x have a vulnerability where code debugging methods are enabled, which could allow an attacker to remotely execute arbitrary code during runtime.
CVE-2018-10599 1 Philips 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more 2023-12-10 2.9 LOW 5.3 MEDIUM
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that allows an unauthenticated attacker to read memory from an attacker-chosen device address within the same subnet.
CVE-2018-5474 1 Philips 1 Intellispace Portal 2023-12-10 7.5 HIGH 9.8 CRITICAL
Philips Intellispace Portal all versions 7.0.x and 8.0.x have an input validation vulnerability that could allow a remote attacker to execute arbitrary code or cause the application to crash.
CVE-2018-5462 1 Philips 1 Intellispace Portal 2023-12-10 5.0 MEDIUM 7.5 HIGH
Philips IntelliSpace Portal all versions of 8.0.x, and 7.0.x have an SSL incorrect hostname certificate vulnerability this could allow an attacker to gain unauthorized access to resources and information.
CVE-2017-9658 1 Philips 2 Intellivue Mx40, Intellivue Mx40 Firmware 2023-12-10 6.1 MEDIUM 6.5 MEDIUM
Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
CVE-2018-7498 1 Philips 2 Alice 6, Alice 6 Firmware 2023-12-10 5.0 MEDIUM 9.8 CRITICAL
In Philips Alice 6 System version R8.0.2 or prior, the lack of proper data encryption passes up the guarantees of confidentiality, integrity, and accountability that properly implemented encryption conveys.
CVE-2018-10601 1 Philips 36 Avalon Fetal\/maternal Monitors Fm20, Avalon Fetal\/maternal Monitors Fm20 Firmware, Avalon Fetal\/maternal Monitors Fm30 and 33 more 2023-12-10 5.4 MEDIUM 8.2 HIGH
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
CVE-2018-5464 1 Philips 1 Intellispace Portal 2023-12-10 5.0 MEDIUM 7.5 HIGH
Philips IntelliSpace Portal all versions of 8.0.x, and 7.0.x have an untrusted SSL certificate vulnerability this could allow an attacker to gain unauthorized access to resources and information.
CVE-2017-14797 1 Philips 2 Hue Bridge Bsb002, Hue Bridge Bsb002 Firmware 2023-12-10 7.9 HIGH 7.5 HIGH
Lack of Transport Encryption in the public API in Philips Hue Bridge BSB002 SW 1707040932 allows remote attackers to read API keys (and consequently bypass the pushlink protection mechanism, and obtain complete control of the connected accessories) by leveraging the ability to sniff HTTP traffic on the local intranet network.
CVE-2017-14111 1 Philips 2 Intellispace Cardiovascular, Xcelera 2023-12-10 4.0 MEDIUM 7.2 HIGH
The workstation logging function in Philips IntelliSpace Cardiovascular (ISCV) 2.3.0 and earlier and Xcelera R4.1L1 and earlier records domain authentication credentials, which if accessed allows an attacker to use credentials to access the application, or other user entitlements.
CVE-2015-2882 1 Philips 1 In.sight B120\\37 2023-12-10 10.0 HIGH 9.8 CRITICAL
Philips In.Sight B120/37 has a password of b120root for the backdoor root account, a password of /ADMIN/ for the backdoor admin account, a password of merlin for the backdoor mg3500 account, a password of M100-4674448 for the backdoor user account, and a password of M100-4674448 for the backdoor admin account.
CVE-2015-2883 1 Philips 1 In.sight B120\\37 2023-12-10 3.5 LOW 5.4 MEDIUM
Philips In.Sight B120/37 has XSS, related to the Weaved cloud web service, as demonstrated by the name parameter to deviceSettings.php or shareDevice.php.
CVE-2015-2884 1 Philips 1 In.sight B120\\37 2023-12-10 5.0 MEDIUM 7.5 HIGH
Philips In.Sight B120/37 allows remote attackers to obtain sensitive information via a direct request, related to yoics.net URLs, stream.m3u8 URIs, and cam_service_enable.cgi.