Vulnerabilities (CVE)

Filtered by vendor Roche Subscribe
Filtered by product Coaguchek Pro Ii
Total 3 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2018-18563 1 Roche 10 Accu-chek Inform Ii, Accu-chek Inform Ii Firmware, Coaguchek Pro Ii and 7 more 2023-12-10 8.3 HIGH 9.6 CRITICAL
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial Number below KQ0400000 or KS0400000) and cobas h 232 before 04.00.04 (Serial Number above KQ0400000 or KS0400000). Improper access control to a service command allows attackers in the adjacent network to execute arbitrary code on the system through a crafted Poct1-A message.
CVE-2018-18565 1 Roche 10 Accu-chek Inform Ii, Accu-chek Inform Ii Firmware, Coaguchek Pro Ii and 7 more 2023-12-10 4.1 MEDIUM 6.8 MEDIUM
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, CoaguChek XS Plus before 03.01.06, CoaguChek XS Pro before 03.01.06, cobas h 232 before 03.01.03 (Serial number below KQ0400000 or KS0400000), and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). A vulnerability in the software update mechanism allows authenticated attackers in the adjacent network to overwrite arbitrary files on the system through a crafted update package.
CVE-2018-18564 1 Roche 6 Accu-chek Inform Ii, Accu-chek Inform Ii Firmware, Coaguchek Pro Ii and 3 more 2023-12-10 3.3 LOW 7.4 HIGH
An issue was discovered in Roche Accu-Chek Inform II Instrument before 03.06.00 (Serial number below 14000) and 04.x before 04.03.00 (Serial Number above 14000), CoaguChek Pro II before 04.03.00, and cobas h 232 before 04.00.04 (Serial number above KQ0400000 or KS0400000). Improper access control allows attackers in the adjacent network to change the instrument configuration.