Vulnerabilities (CVE)

Filtered by vendor Gehealthcare Subscribe
Filtered by product Apexpro Telemetry Server Firmware
Total 6 CVE
CVE Vendors Products Updated CVSS v2 CVSS v3
CVE-2020-6963 1 Gehealthcare 12 Apexpro Telemetry Server, Apexpro Telemetry Server Firmware, Carescape Central Station Mai700 and 9 more 2023-12-10 10.0 HIGH 10.0 CRITICAL
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilized hard coded SMB credentials, which may allow an attacker to remotely execute arbitrary code.
CVE-2020-6964 1 Gehealthcare 12 Apexpro Telemetry Server, Apexpro Telemetry Server Firmware, Carescape Central Station Mai700 and 9 more 2023-12-10 5.0 MEDIUM 8.6 HIGH
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
CVE-2020-6965 1 Gehealthcare 18 Apexpro Telemetry Server, Apexpro Telemetry Server Firmware, Carescape B450 Monitor and 15 more 2023-12-10 6.5 MEDIUM 9.9 CRITICAL
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, a vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package.
CVE-2020-6962 1 Gehealthcare 18 Apexpro Telemetry Server, Apexpro Telemetry Server Firmware, Carescape B450 Monitor and 15 more 2023-12-10 10.0 HIGH 10.0 CRITICAL
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X CARESCAPE Central Station (CSCS) Versions 2.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, an input validation vulnerability exists in the web-based system configuration utility that could allow an attacker to obtain arbitrary remote code execution.
CVE-2020-6966 1 Gehealthcare 12 Apexpro Telemetry Server, Apexpro Telemetry Server Firmware, Carescape Central Station Mai700 and 9 more 2023-12-10 10.0 HIGH 10.0 CRITICAL
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.
CVE-2020-6961 1 Gehealthcare 12 Apexpro Telemetry Server, Apexpro Telemetry Server Firmware, Carescape Central Station Mai700 and 9 more 2023-12-10 7.5 HIGH 10.0 CRITICAL
In ApexPro Telemetry Server, Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Telemetry Server Version 4.3, CARESCAPE Central Station (CSCS) Versions 1.X, a vulnerability exists in the affected products that could allow an attacker to obtain access to the SSH private key in configuration files.